Product Details:
|
Weight: | 400-500KG | Color: | Blue Or Customize |
---|---|---|---|
Power: | 220V , 50Hz | Controller Resolution: | ±0.3℃;±2.5%RH |
Function: | Product Stability Test | Keyword: | Fast Change Rate Test Chamber |
Condenser: | Water-cooled Condensing Agent | Volume: | 80L,150L,225L,408L,800L,1000L Or Custom Made |
High Light: | Environment Test Equipment,Environment Chamber |
Rapid Rate Change Aging Test Chamber , Low Pressure High Altitude Test Chamber
What is an ALTITUDE TEST CHAMBER ?
An altitude chamber combines temperature and low-pressure altitude with optional humidity to test basic components or sub-assemblies for a variety of industries, such as defense, aerospace, avionics, automotive, electronics and others.
Here are two common appliations of an altitude test chamber :
Testing of avionics or other aerospace equipment to simulate actual conditions it might experience
Simulation of high-altitude conditions like might be found in mountainous regions and plateau
And such an altitude test chamber is especially important for defense and aerospace applications where pressure and vacuum conditions are even more intense.
ATMARS can provide custom-built altitude chambers for simultaneous testing of temperature, low-pressure altitude and optional humidity conditions.
Our altitude/temperature and altitude/temperature/humidity chambers offer performance flexibility and reliablility for today's mixed-environment testing requirements with an overall temperature range of -70°C to +150°C, optional humidity capabilities of 20% to 98% RH and altitude up to 200,000 feet.
Altitude Chambers Key Specifications:
Workspace Volume: 250L, 500L, 800L, 1000L to walk-in/drive-in sizes
Temperature Range: -70°C to 150°C (at standard atmospheric pressure)
Optional Humidity Range: 20% to 98% RH (optional humidity control when altitude system is off)
Pressure Range: 0.5KPa to 101.325 KPa (standard atmospheric pressure)
Standard Altitude Range: site level to 100,000 feet (30,480m / 8.28 mm Hg / 8.28 Torr)
Optional Altitude Range: site level to 150,000 feet (45,720m / 1.10 mm Hg / 1.10 Torr) site level to 200,000 feet (60,960m / 0.169 mm Hg / 0.169 Torr)
Optional extended ranges are also available to simulate sea level testing when your site is significantly above sea level
Model | ||
800D | ||
Performance | Temperature range | -70℃ ~ +150℃ |
Temperature fluctuation | ±0.5℃(at a steady temperature) | |
Temperature deviation | ≤±2.0℃ | |
Temperature uniformity | ±2.0℃ | |
Humidity range | 20% RH~98% RH (at Atmospheric pressure) | |
Humiditu fluctuation | ±2.0% RH(at Atmospheric pressure) | |
Humidity deviation | ≤±1.0% RH | |
Cool-down rate | 0.7℃~1.0℃/min | |
Pressure range | Atmospheric pressure to 0.5kPa | |
Pressure deviation | ±2kPa,When ≥40kPa, ±0.1kPa,When ≤4kPa ±0.5kPa,When 4kPa~40kPa |
|
Depressurization rate | ≤30min,when between Atmospheric pressure and 1kPa | |
Pressure recover rate | 10kPa/min adjustable | |
Workspace volume | 800L | |
Workspace dimensions (WxDxH)mm |
800x1000x1000 | |
Exterior dimensions (WxDxH)mm |
1250x2400x2300 | |
Material | Interior | Stainless steel (SUS304) |
Exterior | Cold-rolled steel with powder coated surface | |
Seal | Imported silicone seal | |
Insulation | Expanded foam & fiberglass | |
System | Heating | Nichrome resistance wire heater |
Refrigeration | German GEA cascade semi-hermetic compressors Air-cooled or Water-cooled cooling mode for option Zero Ozone Depletion Potential (ODP) refrigerant - R404A / R23 |
|
Cooling mode | Water-cooling | |
Temperature control mode | Balanced Temperature& Humidity Control System(BTHC) | |
Controller | Display Operating mode Memory capacity Control tolerance Temperature sensor input Communication interface |
TFT true color LED touch-screen; Program and setpoint operating mode; Program capacity up to 127 programs, each 999 hours, maxiumum 32,000 cycles; Temperature measuring range: -90.00℃ to 200.00℃, tolerance: ±0.20℃; Supports multiple sensor inputs such as T type, PT100; Comes with USB interface, supports RS-485, LAN and others |
Power requirement | 18.5KW | |
Noise level | ≤70dB (A-weighted sound pressure level measured at a distance of 1.0 meter) | |
Allowable ambient conditions | Temperature: +5℃ to +35℃; Relative humidity: ≤85%RH (No condensation); Air pressure: 86kPa to 106kPa |
|
Electrical supply | AC 380 (1±10%)V (50±0.5)Hz 3 phase 4 wire + ground wire |
We keep good quality and competitive price to ensure our customers benefit ; We respect every customer as our friend and we sincerely do business and make friends with them,no matter where they come from. Each batch of parts is done quality sampling inspection before into the warehouse, and Each product before delivery is tested by the national authoritative third-party inspection agency
- A variety of industrial and laboratory drying ovens;
- Muffle and chamber furnaces;
Contact Person: Ms. Kris Zhang
Tel: 0086-0769-85914911